REMI Walk in chambers are designed for simulating, monitoring & control of environmental conditions like temperature & humidity. All these chambers are double walled & modular in construction, easy to assemble at site. They are available in standard sizes & can also be tailor made to suit specific requirements. Forced air circulation with the help of fan blower maintains uniform conditions. Use of intelligent control systems & dedicated REMI ‘Datasoft’ software complying to 21 CFR Part 11 requirements greatly enhances the performance. These chambers can be provided with standby refrigeration & humidity systems (optional) for uninterrupted functioning.

CONSTRUCTION
  • Double walled prefabricated modular panels with PUF insulation, easy to assemble at site
  • Standard model with inner of Stainless Steel (304 Grade) and outer Powder coated steel sheet
  • GMP model with inner (316Grade) and outer of Stainless Steel (304 Grade)
  • CFC free Refrigeration system with split type condensing unit located outside lab helps to emit hot air beyond working area
  • Heating by long lasting SS tubular heaters
  • Humidity by steam injection. Boiler and water reservoir made of SS with heater and low water level cut off
  • Forced Air Circulation by heavy duty motor
  • SS racks and trays
Salient Features:
  • Designed as per ICH guidelines to meet, WHO and USFDArequirement
  • Fulfills storage conditions of 25°C-60% RH, 40°C-75% RH, 30°C-65%
  • Temperature Range: 20°C to 60°C, Accuracy:+0.2 C,Uniformity:±1°C (STABILITY CHAMBERS)
  • Humidity range: 40% RH to 95% RH, Accuracy:±2% RH, Uniformity:±3%RH (STABILITY CHAMBERS)
  • Temperature Range:2°C to 8°C, Accuracy:±0.2°C, Uniformity:±1°C (COLD ROOMS)
  • Microprocessor based auto tuning PID controller provided for precise control of temperature and humidity
  • Auto change over to standby systems through PLC (Optional)
  • Auto switch over to standby sensor in case of Controlling Sensor failure through PLC (Optional)
  • Capacitance type Humidity sensor enables direct display of RH and temperature (STABILITY CHAMBERS)
  • Audio visual alarm for temperature, humidity variations and utility failures
  • Output for remote alarm in lab and at security gate
  • Safety includes shut off of humidity and air heaters in case of temperature overshoots or undershoots beyond specified limits with alarms
  • Facility to open the door from inside in case of door locked with emergency bell having switch inside chamber
  • Low water level alarm and power cut off of boiler tank heater
  • Overload cut off relay for compressor
  • Time delay for compressor switch on
  • PC communication via RS-485 and software complying to 21CFR part 11 USFDA guidelines (Optional)
  • Data stored in PC server with RS-485 connectivity and password protection (Optional with software)
  • PLC with touch screen HMI fully protected with password, instead of PID controller for precise control of temperature and humidity (optional)
  • Standby refrigeration system available for uninterrupted functioning (Optional)
  • Standby humidity system available for uninterrupted functioning (Optional)
  • Multipoint Temperature and Humidity logger with printout facility (optional)
  • GSM mobile alert in case of temperature and humidity deviations beyond specified Limits (Optional with software)
  • Magnetic door lock system with biometrics for door access and software to monitor door open log (Optional)
  • Validation protocol with IQ, OQ, PQ Documentation
  1. Intelligent PLC/PID control system helps maintain highly accurate set parameters
  2. Biometrics system for secured door access
  3. Dual Capacitance type Humidity Sensors failure of one sensor automatically transfers regulation to other sensor
  4. Chamber illumination by fluorescent tubes
  5. PUF Insulation 80 mm for maximum thermal protection
  6. Spring Door Latch for door closure
  7. Blower Motor for forced air circulation
  8. Control panel with Data-logger, PC connectivity and electrical components
  9. Standby refrigeration by Compressor-1 and Compressor-2
  10. Standby humidity by Heater-1 and Heater-2
  11. Emergency bell switch to open the door from inside
  12. Highly ergonomic handle with unique multi lever key lock
  13. SS Racks and Trays

LCD Display

Features
  • 4 Inch attractive LCD display
  • Intelligent Controller helps maintain temperature in case of sensor failure
  • Battery Backup for Temperature Controller
  • Auto tuning of controller
  • Self-Diagnosis for errors
  • Histogram format of 24 hours temperature recording
  • Settable High / low alarm points
  • Memory for storage of upto 1000 records
  • Centronic interface to record Temperature, Date, & Time by attaching Dot Matrix printer with adjustable print interval
  • Alarm on equipment / output for remote alarm

Touch Screen Display

Features
  • 7 inch big colour touch screen
  • PLC (Programable logic controller)
  • Memory of 9,000 records
  • Centronic interface to record Temperature, Date & Time by attaching Dot Matrix printer with adjustable print interval
  • Ethernet port
  • Battery backup in case of power failure for display, alarms & off-line data logging (optional)
  • Programmable cyclic timer for controlling illumination conditions
  • Remi “Datasoft” Data acquisition software complying to 21 CFR part 11 as per USFDA Guidelines (optional)
  • Password protected door opening (Optional)

REMI ‘Data Soft’

Features
  • Multiple level alpha numerical password with password ageing
  • Secured audit trail report
  • Alarm report with mention of alarm condition
  • Mean Kinetic Temperature able to be calculated for any days& alarm deviation report
  • Reports in graphical and tabular form
  • Data Stored in PC server with Lan connectivity and can be monitored on PC with password authentication
  • Multi chamber data acquisition on single software
  • Communication through RS 485 / TCP IP
  • GSM Mobile Alert for six Numbers
  • Power failure and resumption recorded with date and time

Optional Accessories

To be Ordered Separately
  • Data logger for 8 Point temperature recording with sensor printer interface and software
  • Software as per 21 CFR Part 11 requirements
  • PLC Based Control System with HMI Touch Screen
  • Biometrics Magnetic Door Access for Secured Door opening with software to maintain log
  • Validation (mapping) consist of one cycle for temperature at any one set point for 24 hours on empty and loaded condition
  • Standby Refrigeration System
  • Standby Humidity System
  • GSM alert through SIM card
Technical Data
WALK-IN STABILITY CHAMBERS WSC -50 WSC -80 WSC -100 WSC -120 WSC -150
WALK-IN COLD ROOM WCR -50 WCR -80 WCR -100 WCR -120 WCR -150
Internal Volume (Liters) 5000 8000 10000 12500 15000
Number of Shelves 16 24 32 48 48
Internal Dimension W x D x H (mm) 1250 x 2000 x 2000 2000 x 2000 x 2000 2000 x 2000 x 2500 2000 x 2500 x 2400 2500 x 2500 x 2400
External Dimension W x D x H (mm) 1410 x 2360 x 2160 2160 x 2360 x 2160 2160 x 2360 x 2660 2160 x 2860 x 2560 2660 x 2860 x 2560
Internal Body Material Stainless Steel 304 grade (Standard Models), Stainless Steel 316 grade (GMP Models)
External Body Material Powder Coated CRCA Steel (Standard Models), Stainless Steel 304 grade (GMP Models)
Insulation 80 mm minimum, CFC free polyurethane foam
Illumination Fluorescent Tubes
Control & Display Microprocessor based PID control system with large 4″ LCD Display
Power Failure Alarm Audio Visual Alarm
Door Open Alarm Audio Alarm in case door open for over one minute

Supply: 220-240 Volts, 50 Hz Single Phase.

Walk-in Stability Chambers Walk-in Cold Rooms
Sensor Capacitance type direct reading PT-100
Temperature Range & Accuracy 20°C to 60°C, ± 1°C 2°C to 8°C, ± 1°C
Humidity Range (Rh) & Accuracy 40% to 95%, ± 3% RH xx
Audio Visual Variation Alarm Set Humidity ± 3% RH,, Set Temperature ± 2°C Set Temperature ± 2°C